Land - System 4 High Precision Thermometers

Manufacturer:
SYSTEM 4 thermometers offer exceptional flexibility with a choice of single wavelength, ratio, fibroptic and fibroptic ratio models.
Part Number: Land System 4 Thermometers
Back to Previous Page
Product Files
Call for pricing

Non-contact, high precision temperature measurement

SYSTEM 4 thermometers offer exceptional flexibility with a choice of single wavelength, ratio, fibroptic and fibroptic ratio models.

Thermometer type, temperature range, spectral response and optical characteristics are chosen to suit any application from 0 to 2600°C.

Key Features

  • Fully interchangeable thermometer and processors ensure maximum flexibility without the need for recalibration.
  • Wide range of thermometers with focusable, through-the-lens sighting or flexible fibroptics with optional laser targeting.
  • Small target capability
  • Wide range of highly effective mounting accessories
  • Rugged, flexible, modular design
  • Accurate, reliable, drift-free measurement
  • Industry standard outputs to suit any process monitoring, recording or control system

Details

SYSTEM 4 comprises an advanced range of high precision radiation thermometers, LANDMARK® processors and a range of mounting accessories which combine to form a complete temperature measurement system.

Thermometer type, temperature range, spectral response and optical characteristics are chosen to suit any application from 0 to 2600°C/ 32 to 4700°F.

  • Focusable optics - standard and short focus versions with through-the-lens sighting providing clear and guaranteed definition of target.
  • Optional close-up lenses - giving measurement of targets as small as 0.45mm.
  • Accurate, reliable, drift-free measurement.
  • Rugged design with a range of mounting options.
  • Flexible fibre optics light guide versions - with optional laser targeting system to define target spot.
  • High level linear output.

Radiation Thermometers

Proven, reliable electronics and a high quality optical system combine to give a thermometer which delivers accurate, dependable temperature measurement. A rugged die-cast aluminium body, with a high quality electrical connector, ensures reliable performance.

Standard bodied thermometers all feature through-the-lens sighting with a 6° field of view. Adjustable focus with a circular graticule gives precise alignment on to the smallest of targets. Two optical variants are available: Standard focus - adjustable between 500mm and infinity, and Short-focus - viewing from 350mm to 1m. Close-up lenses are also available which can measure targets as small as 0.45mm from as close as 90mm.

Fibroptic Thermometers

The use of flexible fibre optics light guides allows the detector and electronics enclosure to be located in a less hostile environment, and enables access to difficult targets.

The fibroptic thermometers are available with an optional integral laser targeting system which defines the target spot for accurate alignment.

The use of fibre optics permits viewing of normally inaccessible targets, where there are high magnetic fields or in high ambient temperatures up to 200 °C without cooling of the optic head. There is a choice of three optic heads and three light guide lengths.

Emissivity Enhancer

The Emissivity Enhancer improves the accuracy of System 4 short wavelength thermometers on difficult to measure materials such as bright annealed and stainless steel strip.

Intended for use on low emssivity materials it gives enhanced emissivity values.

Applications

  • Metals
  • Glass surface and glass processing
  • Refractories
  • Food
  • Paper
  • Textiles
  • Plastics
  • Wood
  • Ceramics
Temperatures
Low Temperature (C): 0 High Temperature (C): 2600
Low Temperature (F): 32 High Temperature (F): 4712

Technical Details

Standard Thermometers

ModelRangeWavelengthResponse Time (1)Accuracy (2)
General purpose, metals, glass, refractories    
M1 450/1000C 450 to 1000 °C 1µm 5 ms 0.4%K
M1 850/1850F 850 to 1850 °F 1µm 5 ms 0.4%K
M1 600/1600C 600 to 1600 °C 1µm 5 ms 0.4%K
M1 1100/2900F 1100 to 2900 °F 1µm 5 ms 0.4%K
M1 800/2600C 800 to 2600 °C 1µm 5 ms 0.7%K
M1 1500/4700F 1500 to 4700 °F 1µm 5 ms 0.7%K
General purpose, metals and glass processing    
M2 300/1100C 300 to 1100 °C 1.6 µm 5 ms 0.25% + 1K
M2 600/2000F 600 to 2000 °F 1.6 µm 5 ms 0.25% + 1K
Bright oxidised or lightly oxidised metal surfaces    
M4 50/250C 50 to 250 °C 2.4 µm 100 ms 3K
M4 150/500F 150 to 500 °F 2.4 µm 100 ms 3K
M4 150/550C 150 to 550 °C 2.4 µm 100 ms 4K
M4 300/1000F 300 to 1000 °F 2.4 µm 100 ms 4K
Glass surface temperature measurement    
M5 400/1300C 400 to 1300 °C 4.8 to 5.2 µm 100 ms 0.6%K (4)
M5 750/2400F 750 to 2400 °F 4.8 to 5.2 µm 100 ms 0.6%K (4)
M5 1000/2500C 1000 to 2500 °C 4.8 to 5.2 µm 100 ms 0.6%K (4)
M5 1800/4500F 1800 to 4500 °F 4.8 to 5.2 µm 100 ms 0.6%K (4)
General purpose, low temperature, metals, glass, food, paper and textiles
M6 0/300C 0 to 300 °C 3 to 5 µm 100 ms 0.3% + 2.5K
M6 50/600F 50 to 600 °F 3 to 5 µm 100 ms 0.3% + 2.5K
M6 100/700C 100 to 700 °C 3 to 5 µm 100 ms 0.3% + 2K
M6 200/1300F 200 to 1300 °F 3 to 5 µm 100 ms 0.3% + 2K
Plastic thin film    
M7 25/375C 25 to 375 °C 3.43 µm 100 ms 3K
M7 75/700F 75 to 700 °F 3.43 µm 100 ms 3K
Food, textiles, paper, plastics, rubber, wood, paint and ceramics
M8 0/1000C 0 to 1000 °C 8 to 14 µm 100 ms 1%K + 1K
M8 30/1830F 30 to 1830 °F 8 to 14 µm 100 ms 1%K + 1K
Ratio thermometer, variable emissivity, target obscuration, partially filled field of view
R1 600/1600C 600 to 1600 °C 0.85 to 1.1 µm 100 ms 0.65%K
R1 1100/2900F 1100 to 2900 °F 0.85 to 1.1 µm 100 ms 0.65%K
R1 1000/2600C 1000 to 2600 °C 0.85 to 1.1 µm 100 ms 1.1%K
R1 1800/4700F 1800 to 4700 °F 0.85 to 1.1 µm 100 ms 1.1%K

(1) - Time quoted to 95% step change
(2) - Accuracy quoted to ITS90
(3) - Above 75 °C / 167 °F
(4) - Optimised for glass toughening = 3k at 630 °C

Fibroptic Thermometers

ModelRangeWavelengthResponse Time (1)Accuracy (2)
General purpose, metals, glass, refractories    
M1 600/1600CYL 600 to 1600 °C 1µm 5 ms 0.4%K
M1 1100/2900FYL 1100 to 2900 °F 1µm 5 ms 0.4%K
M1 800/2600CYL 800 to 2600 °C 1µm 5 ms 0.7%K
M1 1500/4700FYL 1500 to 4700 °F 1µm 5 ms 0.7%K
General purpose, metals and glass processing    
M2 300/1100CYL 300 to 1100 °C 1.6 µm 5 ms 0.25% + 1K
M2 600/2000FYL 600 to 2000 °F 1.6 µm 5 ms 0.25% + 1K
General purpose, low temperature and secondary metals    
M3 50/250CQ (3) 50 to 250 °C 2.1 µm 100 ms 3K
M3 150/500FQ (3) 150 to 500 °F 2.1 µm 100 ms 3K
Ratio thermometer, variable emissivity, target obscuration, partially filled field of view
R1 600/1600CYL 600 to 1600 °C 0.85 to 1.1 µm 100 ms 0.65%K
R1 1100/2900FYL 1100 to 2900 °F 0.85 to 1.1 µm 100 ms 0.65%K
R1 1000/2600CYL 1000 to 2600 °C 0.85 to 1.1 µm 100 ms 1.1%K
R1 1800/4700FYL 1800 to 4700 °F 0.85 to 1.1 µm 100 ms 1.1%K

Y - Denotes laser targetting system available
(3) - Above 75 °C / 167 °F

Land Applications & Industries


Cement

LAND have supplied the Cement Industry and OEM's throughout the world with temperature measuring equipment for over 60 years. All their products are designed to the highest standards of quality and reliability to ensure accurate measurements under plant operating conditions for; kiln shell, rotary kiln product & burning zone and clinker system read more >>


Iron and Steel Making

Designed to operate in the harsh environments found in steel production, these instruments are designed to the highest performance standards, optimised for making temperature measurements at every important stage of the process including; Soaking Pit (Load Temperature), Coke Oven Flue, profile, burden & conveyor, Hot blast stove, main & furnace burden, liquid iron stream, torpedo car,  ladle pre heat, ladle and tapping learn more >>


Aluminium Extrusion

LAND offers a complete new integrated range of thermometer systems dedicated solely to aluminum temperature measurement. These instruments contain unique-to-Land design features which provide inherent measurement stability and reliability. In the aluminum extrusion industry, product quality and operating profitability rely on efficient press operation. Press efficiency is increased greatly with accurate, reliable temperature measurement at critical points in the process including; preheated die, billet profile , die exit & quench exit temperatures read more >>


Aluminium Rolling

The Land Aluminium Strip Thermometer system is specifically designed to provide the aluminium hot rolling industry with a solution to one of its most difficult problems - accurate and reproducible temperature measurement of bright, semi-finished hot alloy strips learn more >>





More Industry Applications


Land has supported the global industrial community for more than 70 years, supplying innovative, reliable products and solutions. They have a worldwide reputation as innovators and manufacturers of high-quality, precision thermometers and thermal imagers.

Land's monitors and analysers provide industries with infrared, non-contact temperature measurement, combustion efficiency monitors and environmental pollution emissions detectors. Their products are used extensively in applications as diverse as steel, glass, electronics, mineral processing, power generation, utility and aircraft gas turbines, and a wide range of other industries. Ongoing development creates products for new applications and the continuous improvement of an expanding product range.

  • Steel: The infrared thermometers offer process solutions across more than 50 individual applications, ensuring process efficiency and continue reading >>
  • Aluminium: Including a wide range of stable, rapid response and reliable application-specific instruments designed to the highest standards continue reading >>
  • Speciality Metals: Products that you need for precise temperature measurements during speciality metal production, including solutions for copper and platinum refining continue reading >>
  • Calibration: The blackbody calibration sources provide a high-precision calibration of radiation thermometers, on-site and under laboratory conditions continue reading >>
  • Glass: Land has developed a range of instruments specifically designed to meet the challenges of glass production continue reading >>
  • Hydrocarbon Processing (HPI): Built to the highest standards of quality and reliability, their HPI measurement solutions are designed to operate in industrial conditions, delivering the accuracy you need to control your process continue reading >>
  • Industrial Processing: Used extensively in a range of industrial processes including paper, plastics, ceramics and non-wovens, Land's accurate online and portable thermometers, linescanners and thermal imagers support your applications continue reading >>
  • Minerals: Land supplies high-precision, reliable non-contact temperature measurement instruments specifically designed for mineral industry applications, including cement and roadstone production continue reading >>
  • Power Generation: With a range of detection options for every stage of the combustion process, Land's products guard against spontaneous fires. They also provide solutions for continuous emissions monitoring and plant condition monitoring continue reading >>

Non-contact, high precision temperature measurement

SYSTEM 4 thermometers offer exceptional flexibility with a choice of single wavelength, ratio, fibroptic and fibroptic ratio models.

Thermometer type, temperature range, spectral response and optical characteristics are chosen to suit any application from 0 to 2600°C.

Key Features

  • Fully interchangeable thermometer and processors ensure maximum flexibility without the need for recalibration.
  • Wide range of thermometers with focusable, through-the-lens sighting or flexible fibroptics with optional laser targeting.
  • Small target capability
  • Wide range of highly effective mounting accessories
  • Rugged, flexible, modular design
  • Accurate, reliable, drift-free measurement
  • Industry standard outputs to suit any process monitoring, recording or control system

Details

SYSTEM 4 comprises an advanced range of high precision radiation thermometers, LANDMARK® processors and a range of mounting accessories which combine to form a complete temperature measurement system.

Thermometer type, temperature range, spectral response and optical characteristics are chosen to suit any application from 0 to 2600°C/ 32 to 4700°F.

  • Focusable optics - standard and short focus versions with through-the-lens sighting providing clear and guaranteed definition of target.
  • Optional close-up lenses - giving measurement of targets as small as 0.45mm.
  • Accurate, reliable, drift-free measurement.
  • Rugged design with a range of mounting options.
  • Flexible fibre optics light guide versions - with optional laser targeting system to define target spot.
  • High level linear output.

Radiation Thermometers

Proven, reliable electronics and a high quality optical system combine to give a thermometer which delivers accurate, dependable temperature measurement. A rugged die-cast aluminium body, with a high quality electrical connector, ensures reliable performance.

Standard bodied thermometers all feature through-the-lens sighting with a 6° field of view. Adjustable focus with a circular graticule gives precise alignment on to the smallest of targets. Two optical variants are available: Standard focus - adjustable between 500mm and infinity, and Short-focus - viewing from 350mm to 1m. Close-up lenses are also available which can measure targets as small as 0.45mm from as close as 90mm.

Fibroptic Thermometers

The use of flexible fibre optics light guides allows the detector and electronics enclosure to be located in a less hostile environment, and enables access to difficult targets.

The fibroptic thermometers are available with an optional integral laser targeting system which defines the target spot for accurate alignment.

The use of fibre optics permits viewing of normally inaccessible targets, where there are high magnetic fields or in high ambient temperatures up to 200 °C without cooling of the optic head. There is a choice of three optic heads and three light guide lengths.

Emissivity Enhancer

The Emissivity Enhancer improves the accuracy of System 4 short wavelength thermometers on difficult to measure materials such as bright annealed and stainless steel strip.

Intended for use on low emssivity materials it gives enhanced emissivity values.

Applications

  • Metals
  • Glass surface and glass processing
  • Refractories
  • Food
  • Paper
  • Textiles
  • Plastics
  • Wood
  • Ceramics