InTemp CX400 Series Comparison
|
|
|
|
|
|
|
|
Who is this for? |
Required to follow GDP & GMP requirements specifically
- Probe in glycol
- GDP/GMP compliant
- Frozen to ambient monitoring
|
Required to follow VFC & CDC requirements specifically
- Probe in glycol
- CDC/VFC compliant with 3 year NIST certificate
- GDP/GMP compliant
- Frozen to refrigerated monitoring
|
Required to follow GDP & GMP requirements specifically
- Probe only (no glycol bottle)
- GDP/GMP compliant
- Frozen to ambient monitoring
|
Required to follow GDP & GMP requirements specifically
- Internal sensor only (No probe)
- GDP/GMP compliant
- Refrigerated to ambient monitoring
|
Required to follow GDP & GMP requirements specifically
- Ultra-low RTD probe (no glycol bottle)
- GDP/GMP compliant
- Ultra-low dry ice to refrigerated monitoring
|
Required to follow GDP & GMP requirements specifically
- Internal temp and humidity sensor only (No probe)
- GDP/GMP compliant
- Refrigerated to ambient monitoring
- Humidity monitoring
|
What does it monitor? |
Temperature with probe and glycol bottle and ambient room temperature |
Temperature with probe and glycol bottle and ambient room temperature |
Temperature with probe only and ambient room temperature
|
Ambient Room Temperature |
Ultra low temperature with probe only and ambient room temperature |
Ambient Temperature and Relative Humidity |
NIST Calibration |
One point NIST calibration - Probe and internal ambient sensor |
One point NIST calibration - probe only |
One point NIST calibration - Probe and internal ambient sensor |
One point NIST calibration - Internal ambient sensor only |
Two point NIST calibration - Probe only |
Two point NIST calibration - Internal ambient sensor only, temp & humidity (2 points each) |
NIST Expiration |
1 year NIST certificate - As recommended by GDP/GMP |
3 year NIST certificate - As recommended by VFC/CDC |
1 year NIST certificate - As recommended by GDP/GMP |
1 year NIST certificate - As recommended by GDP/GMP |
1 year NIST certificate - As recommended by GDP/GMP |
1 year NIST certificate - As recommended by GDP/GMP |
NIST Points |
Probe: 5.0°C Internal Sensor: 25.0°C |
Probe: 5.0°C |
Probe: 5.0°C Internal Sensor: 25.0°C |
Internal Sensor: 25.0°C |
Probe Point 1: -70.0°C Probe Point 2: 5.0°C |
Int. Sensor Point 1: 0°C Int. Sensor Point 2: 40.0°C Humidity Point 1: 35.0% @ 25.0°C Humidity Point 2: 75.0% @ 25.0°C
|
External Temperature Probe
|
Range |
-40° to 100°C |
-40° to 100°C |
-40° to 100°C |
N/A |
-95° to 50°C |
N/A |
Accuracy |
±1.0°C from -40° to -15°C ±0.5°C from -15° to 80°C ±0.7°C from 80° to 100°C |
±1.0°C from -40° to -22°C ±0.5°C from -22° to 50°C ±1.0°C from 50° to 100°C |
±0.5°C from -15° to 80°C ±1.0°C from -40° to -15°C ±0.7°C from 80° to 100°C |
N/A |
±1.5°C from -95° to -40°C ±1.0°C from -40° to -10°C ±0.5°C from -10° to 50°C |
N/A |
Resolution |
0.024°C at 25°C |
0.024°C at 25°C |
0.024°C at 25°C |
N/A |
0.05°C at 25°C |
N/A |
NIST Calibration |
Single point NIST calibration, probe and logger body
|
Single point NIST calibration, probe only |
Single point NIST calibration, probe and logger body |
N/A |
Two-point NIST calibration, probe only |
N/A |
Cable Length |
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable
|
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable |
2 or 4 meter (6.56 or 13.12 feet) flat ribbon cable |
N/A |
1.5 m |
N/A |
Probe Dimensions |
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter |
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter |
Food-grade stainless steel probe with pointed tip, 53.34 mm (2.1 inches) long, 3.18 mm (0.125 inches) diameter |
N/A |
Food-grade stainless steel probe with blunt tip, 31.75 mm (1.25 inches) long, 4.78 mm (0.188 inches) diameter |
N/A |
Temperature Sensor
|
Range |
-30° to 70°C |
-30° to 70°C` |
-30° to 70°C |
-30° to 70°C |
-30° to 70°C |
-30° to 70°C |
Accuracy |
±0.5°C from -15° to 70°C ±1.0°C from -30° to -15°C |
±1.0°C from -30° to -22°C ±0.5°C from -22° to 50°C ±1.0°C from 50° to 70°C |
±0.5°C from -15° to 70°C ±1.0°C from -30° to -15°C |
±0.5°C from -15° to 70°C ±1.0°C from -30° to -15°C |
±0.5°C from -15° to 70°C ±1.0°C from -30° to -15°C |
±0.5°C from -30° to 70°C |
Resolution |
0.024°C at 25°C |
0.024°C at 25°C |
0.024°C at 25°C |
0.024°C at 25°C |
0.024°C at 25°C |
0.024°C at 25°C |
RH Sensor
|
Range |
N/A |
N/A |
N/A |
N/A |
N/A |
|
Accuracy |
N/A |
N/A |
N/A |
N/A |
N/A |
|
Resolution |
N/A |
N/A |
N/A |
N/A |
N/A |
|
Logger
|
Logger Operating Range |
-30° to 70°C |
-30° to 70°C |
-30° to 70°C |
-30° to 70°C |
-30° to 70°C |
-30° to 70°C |
Environmental Rating |
IP54 |
IP54 |
IP54 |
IP54 |
IP54 |
IP54 |
Logging Rate |
1 second to 18 hours |
1 second to 18 hours |
1 second to 18 hours |
1 second to 18 hours |
1 second to 18 hours |
1 second to 18 hours |
Transmission Range |
Approximately 30.5 m (100 ft) line-of-sight |
Approximately 30.5 m (100 ft) line-of-sight |
Approximately 30.5 m (100 ft) line-of-sight |
Approximately 30.5 m (100 ft) line-of-sight |
Approximately 30.5 m (100 ft) line-of-sight |
Approximately 30.5 m (100 ft) line-of-sight |
Memory |
128 KB (84,650 measurements, maximum) |
128 KB (84,650 measurements, maximum) |
128 KB (84,650 measurements, maximum) |
128 KB (84,650 measurements, maximum) |
128 KB (75,245 total measurements or 37,622 probe & internal paired measurements) |
128 KB (75,245 total measurements or 37,622 probe & internal paired measurements) |
Full Memory Download Time |
Approximately 60 seconds; may take longer the farther the device is from the logger |
Approximately 60 seconds; may take longer the farther the device is from the logger |
Approximately 60 seconds; may take longer the farther the device is from the logger |
Approximately 60 seconds; may take longer the farther the device is from the logger |
Approximately 60 seconds; may take longer the farther the device is from the logger |
Approximately 60 seconds; may take longer the farther the device is from the logger |
LCD |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
Battery Type |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
LCD is visible from 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range |
Battery Life |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
1 year, typical with logging interval of 1 minute. Faster logging intervals, remaining connected with the InTemp app, excessive report generation, numerous audible alarms, and paging all impact battery life. |
Wireless Data Standard |
Bluetooth Low Energy (Bluetooth Smart) |
Bluetooth Low Energy (Bluetooth Smart) |
Bluetooth Low Energy (Bluetooth Smart) |
Bluetooth Low Energy (Bluetooth Smart) |
Bluetooth Low Energy (Bluetooth Smart) |
Bluetooth Low Energy (Bluetooth Smart) |
Radio Power |
1 mW |
1 mW |
1 mW |
1 mW |
1 mW |
1 mW |
Dimensions |
9.4 x 4.5 x 2.59 cm |
9.4 x 4.5 x 2.59 cm |
9.4 x 4.5 x 2.59 cm |
9.4 x 4.5 x 2.59 cm |
9.4 x 4.5 x 2.59 cm |
9.4 x 5.6 x 2.59 cm |