Teachnical Information
|
Case dimensions: |
360 x 270 x 150 mm |
Weight: |
4.5 kg |
Material: |
Diecast, front foil polyester, ABS |
Sensor inputs: |
4/8/12 sensor inputs for analogueue and digital sensors; freely allocatable. See options Digital CS sensors for dew point and flow with SDI interface FA/VA series, digital third-party sensors RS485 / Modbus RTU. Analogue CS Sensors for pressure, temperature, clamp-on ammeters preconfigured. Analogue third-party sensors 0/4...20 mA, 0...1/10/30 V, pulse, Pt 100 / Pt 1000, KTY, counter |
Voltage supply for sensor: |
24 VDC, max. 130 mA per sensor, integrated mains unit, max. 24 VDC, 25 W. For version 8/12 sensor inputs 2 integrated mains units, each max. 24 VDC, 25 W |
Interfaces: |
USB stick, Ethernet / RS 485 Modbus RTU / TCP, SDI other bus systems on request, webserver optionally, GSM module |
Memory Card: |
Memory size 16 GB Micro SD memory card |
Power supply: |
100…240 VAC, 50-60 Hz |
Colour screen: |
7" touch panel TFT transmissive, graphics, curves, statistics |
Accuracy: |
Please see sensor specifications |
Operating temperature: |
0…50 °C |
Storage temperature: |
-20…70 °C |
Input Signals
|
Current signal internal or external power supply |
(0…20 mA/4…20 mA) |
Measuring range |
0…20 mA |
Resolution |
0.0001 mA |
Accuracy |
0001 mA ± 0.03 mA ± 0.05 % |
Input resistance |
50 Ω |
Voltage signal |
(0…1 V) |
Measuring range |
0…1 V |
Resolution |
0.05 mV |
Accuracy |
± 0.2 mV ± 0.05 % |
Input resistance |
100 kΩ |
Voltage signal |
(0...10 V / 30 V) |
Measuring range |
0…10 V |
Resolution |
0.5 mV |
Accuracy |
± 2 mV ± 0.05 % |
Input resistance |
1 MΩ |
RTD Pt 100 |
|
Measuring range |
-200…850 °C |
Resolution |
0.1 °C |
Accuracy |
± 0.2 °C (-100…400 °C) ± 0.3 °C (further range) |
RTD Pt 1000 |
|
Measuring range |
-200…850 °C |
Resolution |
0.1 °C |
Accuracy |
± 0.2° (-100…400 °C) |
Pulse |
|
Measuring range |
Min pulse length 100 µs frequency 0...1 kHz max. 30 VDC |