Technical Information
|
Measurement Specifications
|
Temperature Range |
-G models: 100°C to 1200°C -GH models: 100°C to 2400°C
Measurements below 100°C are possible with reduced stability
|
Analogue Output Scale |
Factory set: -G models: 4 mA = 50°C; 20 mA = 1200°C -GH models: 4 mA = 50°C; 20 mA = 2400°C Adjustable via optional touch screen module or RS232 |
Response Time (95%) |
-G models: 50 ms -GH models: 10 ms Adjustable up to 5 s via averaging function |
Measured Spot Diameter |
From 2.2 mm depending on measurement distance and choice of optics (see Field of View Diagrams) |
Target Sighting |
Red LED built-in as standard on all models, shows the position and size of the measurement area. Switchable on/off. |
Accuracy of Measurement |
All models: ± 3°C or 1%, whichever is greater -GH models: ± 2% above 1200°C |
Repeatability |
-G models: ± 1°C -GH models: ± 0.2% + 2°C |
Temperature Resolution |
0.5°C |
Emissivity Setting |
Adjustable, 0.3 to 1.0, via RS232C or optional touch screen interface |
Electrical Specifications
|
Outputs |
1 analogue output and 1 alarm output |
Analogue Output |
4-20 mA (this is the default factory setting on all models)
Output type can be changed via optional PM030 touch screen interface to a choice of: 0-20 mA mV/°C Voltage
|
Alarm Output |
1 open drain alarm output, rated 27 V DC, 0.2 A |
Supply Voltage |
5 to 27 V DC, 100 mA max |
Digital Communications |
RS232C Modbus RTU, non-isolated |
Output Cable Connection to Sensor |
-G models: Hardwired -GH models: Connector built into sensor |
Mechanical Specifications
|
Weight (Without Cable) |
-G models: 85 g -GH models: 190 g |
Environmental Specifications
|
Environmental Rating |
IP67 |
Operating Ambient Temperature |
0°C to 50°C |
Storage Temperature |
-15°C to 70°C |
Operating Ambient Humidity |
30% to 85% RH non-condensing |
Display Options
|
Model PM030 |
Optional touch screen terminal for indication, configuration, data logging and alarm outputs. Sensor is available with or without display. |